상세 보기
초록
Complete blood cell count(CBC) is a technique that counts leukocytes for each type of blood cell being analyzed. The principle is that laser is incident to ex vivo flowing leukocytes in a microcapillary tube and scattered light occurs by laser and leukocytes. By collecting the scattered light, we can count the types of cells because different cells generate different light-scattering patterns. However, the technique has an intrinsic limitation, scattering pattern is shown in a wide range region in the resulting, which makes it difficult to accurate analyze and use fluorescent dyes. To overcome this limitation, a new design of CBC with a dual laser, which irradiates with orthogonal angles for collecting quad-scattering information was proposed. Before development, the scattering difference depending on wavelength must be investigated to only catch up to the scattered signal by angles. Some studies, which focused on simple particles, have been conducted to theoretically and experimentally investigate different scatterings by wavelength. In this study, we propose an optical system for measuring scattered light and investigate a complex particle. As a result, the green laser made strong scattering signals in both the forward and side direction: 10% and 30%, respectively.
키워드
- 제목
- 2개의 레이저 파장에 따른 마우스 호중구의 산란광 비교 연구
- 제목 (타언어)
- Comparison of Scattered Light of ex vivo Mouse Neutrophils by Different Wavelength Laser Irradiation
- 저자
- 박재성; 손민지; 황창순; 이호
- 발행일
- 2022-06
- 유형
- Y
- 저널명
- 한국산업융합학회논문집
- 권
- 25
- 호
- 3
- 페이지
- 365 ~ 378
- 언어
- KOR
- 출판사
- 한국산업융합학회
- 발행국가
- 대한민국
- 분량
- 14 페이지
- ISSN
- E 2765-5415
P 1226-833x