상세 보기
자기공명 호환성 향상을 위한 롬버스 메커니즘 기반의 이식형 골전도 압전트랜스듀서 설계
- 신동호;
- 임형규;
- 김명남;
- 성기웅
초록
This study introduces a novel piezoelectric transducer for bone conduction implants that combines piezoelectric elements with a rhombus mechanism to enhance compatibility with magnetic resonance environments. To derive the optimal design of the rhombus structure, various parameters were investigated using theoretical analysis and finite element analysis. A theoretical model of the rhombus structure was employed to identify the parameters affecting displacement amplification magnitude. Based on this, a parametric analysis was performed to calculate the displacement amplification ratio according to these parameters. The results showed that as the beam thickness and width increased, the amplification ratio reduced, while with an increased length, the amplification ratio was increased. Therefore, the optimal rhombus structure for the transducer featured beam dimensions of 0.15 mm thickness, 2 mm width, 3.5 mm length, and 5.5° inclination. This configuration amplified piezoelectric element displacement by a factor of 7.02. The amplification ratio remained constant as long as the mass applied to the rhombus frame to control mechanical resonance did not exceed the blocking force of the piezoelectric element. When a mass of 0.3 g was applied to the frame, mechanical resonance occurred at a frequency of 2 kHz, making it suitable as a transducer for bone conduction implants.
키워드
- 제목
- 자기공명 호환성 향상을 위한 롬버스 메커니즘 기반의 이식형 골전도 압전트랜스듀서 설계
- 제목 (타언어)
- Design of Bone Conduction Implants Piezoelectric Transducer Based on Rhombus Mechanism for Magnetic Resonance Compatibility Improvement
- 저자
- 신동호; 임형규; 김명남; 성기웅
- 발행일
- 2023-08
- 유형
- Y
- 저널명
- 멀티미디어학회논문지
- 권
- 26
- 호
- 8
- 페이지
- 956 ~ 964
- 언어
- KOR
- 출판사
- 한국멀티미디어학회
- 발행국가
- 대한민국
- 분량
- 9 페이지
- ISSN
- P 1229-7771