Nonlinear rheology of stress-controlled rheometers: Large amplitude oscillatory shear

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

Although stress-controlled rheometers are more popular than strain-controlled rheometers, the latter is preferred to the former. This is because stress-controlled rheometers are apt to suffer from inertia effects. This work demonstrates that stress-controlled rheometers can also be used in the same manner as strain-controlled rheometers to study large amplitude oscillatory shear (LAOS). We demonstrate that the inertia effect can be detoured by employing a sufficiently long conditioning time and that the stress decomposition theory remains valid for stress-controlled LAOS data even when strain is not sinusoidal. However, the decomposed stresses are hard to say elastic and viscous because non-sinusoidal strain and its time derivative do not satisfy gamma(2)(t) + omega(-2)(d gamma/dt)(2 )= gamma(2)(o), where gamma(o) is a positive constant. We are certain that cost-effective stress-controlled rheometers can be used if the purpose of LAOS utilization is to identify the fingerprint of rheologically complex materials.

키워드

CONSTITUTIVE-EQUATIONS; BEHAVIOR; INERTIA
제목
Nonlinear rheology of stress-controlled rheometers: Large amplitude oscillatory shear
저자
Choi, Jehyeok; Kim, Min Chan; Hyun, Kyu; Cho, Kwang Soo
DOI
10.1063/5.0219582
발행일
2024-09
유형
Article
저널명
Physics of Fluids
권
36
호
9