2단 절개 형태를 가지는 악교정 수술 장치 연구

A Study of Orthognathic Surgical Guides with Two-stage Split Path
  • 김민욱; 
  • 박충환; 
  • 노지형; 
  • 정의성; 
  • 박영상; 
  • ... 이종민; 
  • 외 3명

초록

In this study, the accuracy of the orthognathic surgical guides with single-stage split path was upgraded to realize orthognathic surgical guides with two-stage split path and simulated surgery was performed to verify its accuracy. As a result, the average error distance between the simulation model and the scan model was + 0.289 / - 0.468 mm (standard deviation 0.128), which was confirmed to be within ± 0.5 mm, which is a clinically acceptable level. Also, there was no significant difference compared with the average value of + 0.313 / - 0.456 mm (average standard deviation 0.106) of the conventional single-stage split path type device. It is judged that the use of this device can contribute to the reduction of surgical time and increase in accuracy since a separate finishing operation for bone preparation is unnecessary.