상세 보기
2단 절개 형태를 가지는 악교정 수술 장치 연구
- 김민욱;
- 박충환;
- 노지형;
- 정의성;
- 박영상;
- ... 이종민;
- 외 3명
초록
In this study, the accuracy of the orthognathic surgical guides with single-stage split path was upgraded to realize orthognathic surgical guides with two-stage split path and simulated surgery was performed to verify its accuracy. As a result, the average error distance between the simulation model and the scan model was + 0.289 / - 0.468 mm (standard deviation 0.128), which was confirmed to be within ± 0.5 mm, which is a clinically acceptable level. Also, there was no significant difference compared with the average value of + 0.313 / - 0.456 mm (average standard deviation 0.106) of the conventional single-stage split path type device. It is judged that the use of this device can contribute to the reduction of surgical time and increase in accuracy since a separate finishing operation for bone preparation is unnecessary.
키워드
- 제목
- 2단 절개 형태를 가지는 악교정 수술 장치 연구
- 제목 (타언어)
- A Study of Orthognathic Surgical Guides with Two-stage Split Path
- 저자
- 김민욱; 박충환; 노지형; 정의성; 박영상; 김동국; 서요한; 우영재; 이종민
- 발행일
- 2022-12
- 유형
- Y
- 저널명
- 의공학회지
- 권
- 43
- 호
- 6
- 페이지
- 382 ~ 389
- 언어
- KOR
- 출판사
- 대한의용생체공학회
- 발행국가
- 대한민국
- 분량
- 8 페이지
- ISSN
- E 2288-9396
P 1229-0807