Development of novel lab-on-a-chip platform for high-throughput radioimmunoassay

  • Kim, Jin-Hee; 
  • Lee, So-Young; 
  • Lee, Seung-Kon
Citations

WEB OF SCIENCE

12
Citations

SCOPUS

16

초록

Radioimmunoassay (RIA) is an extremely specific and a highly sensitive type of immunoassay, but the long incubation time and generation of radioactive wastes limit the use of RIA. To complement these disadvantages of RIA, we suggest an advanced type of RIA based on a lab-on-a-chip (LOC) platform: mu-RIA. We designed a microfluidic chip for RIA and optimized the procedures of mu-RIA analysis, including surface modification, immunoreaction time, and washing. Based on the optimized conditions, we conducted a radioimmunoassay on the mu-RIA platform using a commercial RIA kit. With the mu-RIA, 5 min are adequate for analysis. The amount of reagent consumption is significantly reduced compared with conventional RIA. The standard curve with R-2 = 0.9951 shows that we can quantitatively evaluate the amount of antigen present in unknown samples. We show the applicability of mu-RIA for the analysis of biomolecules and the potential of mu-RIA to be a novel platform for high-throughput analysis.

키워드

Radioimmunoassay; RIA; Radioisotopes; Lab-on-a-chip; Microfluidics; IMMUNORADIOMETRIC ASSAY; IMMOBILIZATION; ANTIGEN
제목
Development of novel lab-on-a-chip platform for high-throughput radioimmunoassay
저자
Kim, Jin-Hee; Lee, So-Young; Lee, Seung-Kon
DOI
10.1016/j.apradiso.2020.109526
발행일
2021-02
유형
Article
저널명
Applied Radiation and Isotopes
권
168